Frequently Asked Questions

Connectivity

Yes, both the Accu‑Chek Instant and the Accu‑Chek Instant S meters connect to the Accu‑Chek Smart Pix system using a standard USB cable. There is no need for an extra dongle. Once connected to a computer, the user then can transfer data, set the meter time and date, and set target ranges.

No, there is no WiFi or NFC connectivity with the meter.

The Accu‑Chek Instant meter implements BLE Pass Key Pairing security, the same technology.

The Accu‑Chek Instant meter has a memory capacity to store at least 720 glucose results, including time, date and test result markers. The Accu‑Chek Instant meter can transfer results via Bluetooth Low Energy (BLE) . As with the Accu‑Chek Instant S meter, the Accu‑Chek Instant meter can also use the micro-USB to connect to a PC. The Accu‑Chek Instant and Accu‑Chek Instant S meters are compatible with the following DM solution: Accu‑Chek Smart Pix software (version 2.2). 

Target range indicator

Being too low is life threatening for people with diabetes. Therefore, red has been chosen to be the color indicating “danger.”

Yes, the target ranges in the meter can be changed using the device settings on the Accu‑Chek Smart Pix system.

The IDF Guidelines are used for the High values range and the ADA Guidelines for the Low values range. In range: 70-160 mg/dL (3.9-8.9 mmol/L), below range: 69-10 mg/dL (3.8-0.6 mmol/L), above range: 161-600 mg/dL (9.0–33.3 mmol/L).

Meters

Yes, except in the US and Canada.

No. As seen with almost all products designed for at-home use, the Accu‑Chek Instant or Accu‑Chek Instant S meters do not have a claim for diagnosis.

“Err” with the error code? During our human factors testing, we discovered that there could be some confusion in misinterpreting an error with an actual test result if the meter is held upside down while viewing the display. We implemented this approach to showing error codes so as to eliminate any risk of misinterpretation.

The Accu‑Chek Instant meters support the AST claim for testing on the forearm, upper arm, below the thumb (thenar) and below the little finger (hypothenar). The Accu‑Chek Instant S meters do not support AST claims at the time of launch. However, additional studies are being conducted to achieve our AST claims post-launch for the Accu‑Chek Instant S meter.

At least 720 readings can be stored inside the meter. However only the last result and the 7, 30 and 90-day averages can be reviewed on the meter. To review all values the meter must be connected to one of the software solutions (e.g. Accu‑Chek Smart Pix or Accu‑Chek Connect).

Accu‑Chek Performa meter—will my patients like it? While smaller, the Accu‑Chek Instant/Accu‑Chek Instant S meters have been proven to include and improve upon the features that are so popular with our current users while maintaining its ease of handling.

Performing a control test lets the user know the meter and test strips are working properly. A control test should be performed when:

  • A new test strip box is opened  
  • The test strip container has been left open
  • The test strips are thought to be damaged
  • Checking the meter and test strips
  • The test strips were stored in extreme temperatures, humidity, or both
  • The meter was dropped
  • The test result does not match how the PwD feels
  • Checking to see if the PwD is performing the test correctly

The Accu‑Chek Instant and Accu‑Chek Instant S meters have a fast test time of less than four seconds.

Test results may be seen in 7, 30 and 90-day averages.

Flagging of test results is not possible on the meter itself. However, if a patient chooses to do so, they can add this information vía a compatible data management app.

No, the USB port for any of the Accu‑Chek meters (Accu‑Chek Instant, Accu‑Chek Instant S) is not water- or dust-resistant.

No, the USB cable does not charge the meter’s batteries. Remove the USB cable after downloading results. In fact, leaving the USB cable connected to the meter and PC after download could draw power from the meter batteries. A blood glucose test cannot be performed while the Accu‑Chek Instant or Accu‑Chek Instant S meter is connected to a PC with a USB cable.

Yes, any time the meter is on, the battery is being consumed whether searching for a paired device or uploading results.

As a general rule, with the connectivity of the meter enabled, the battery life is anticipated to be at least 750 tests. Normal testing without connectivity should produce approximately 1,000 test results for both systems.

The Accu‑Chek Instant meter requires two batteries to power the added value features, such as the backlit display and Bluetooth connectivity. These features are not available with the Accu‑Chek Instant S version, so only one battery is required.

The meters will retain the date and time settings for up to 2 hours.

If replaced within a reasonable time, the memory is not lost when the battery is replaced. If date and time no longer appear on the meter’s display after a is replaced. If date and time no longer appear on the meter’s display after a battery change, a sync with the DMS is required.

Enter our website www.accu-chekcac.com and download the Accu-Chek Smart Pix Software.  Connect the Accu-Chek Instant meter to the computer using a USB cable.  Set the time format and a BG  (blood glucose) target range then click Save.  Afterwards click OK on the confirmation message to complete the configuration.  From the Configure meter menu select Configure date and time. If required, clic OK when synchronzing is complete.

The Accu‑Chek Instant meter makes testing effortless. The features which are not heavily used have been removed from the meter. This makes handling and interactions with the system effortless. Additionally, the Accu‑Chek Instant meter has an intuitive target range indicator that instantly shows PwDs how they are doing with their management.

Accuracy

The FAD-GDH enzyme is a key component of the system’s new test strip chemistry, which works in conjunction with other system components to deliver advanced accuracy, highly reliable performance and reduced interferences.

Yes, the Accu‑Chek Instant and Accu‑Chek Instant S meters comply with the requirements of ISO 15197:2013/EN ISO 15197:2015 (In vitro diagnostic test systems – Requirements for blood glucose monitoring systems for self-testing in managing diabetes mellitus). The evaluation report is available in the launch binder.

The Accu‑Chek Instant and Accu‑Chek Instant S meters are our latest development in accuracy from a pioneer in blood glucose testing technologies. They are two of the most accurate meters in our portfolio.

No, the Accu‑Chek Instant and Accu‑Chek Instant S meters are designed to ensure that the battery power level will not impact results given by the meter.

Yes. In addition to accuracy with capillary blood, the Accu‑Chek Instant and Accu‑Chek Instant S meters may be used with venous, arterial and neonate blood.

The Accu‑Chek Instant system fulfills the ISO 15197:2013/EN ISO 15197:2015 standards and delivers even tighter 10/10 accuracy for reliable results you can trust.

The Accu‑Chek Instant meter is one of the most accurate systems our organization has ever manufactured and meets ISO 15197:2013/EN ISO 15197:2015 standards for a highly accurate system, there are other manufacturers that make a similar accuracy claim.accuracy. It has 10/10 analytical accuracy. While this is

10/10 references an ISO-compliant internal testing protocol. For three different lots of test strips, 95% of BG results must fall within ± 10 mg/dL of reference result at glucose concentrations < 100 mg/dL or within ± 10% at glucose concentrations ≥ 100 mg/dL.

Test Strip

The Accu‑Chek Instant system was designed to assist users to easily conduct a test. Several features were designed for this purpose:

  •  Easy-edge test strip has a wide application area compared to leading brands
  •  Strips are placed electrode down in the vial, so that they can be inserted easily into the meter with minimal handling

In order to achieve the highest accuracy and most robust system claims, a combination of palladium and gold is used for the electrodes. Both of these metals are widely utilized in the electronics and medical industries for surgical instruments and metal contacts.

A combination of palladium and gold is used for the electrodes. Both of these metals are widely utilized in the electronics and medical industries for surgical instruments and metal contacts

There are eight electrodes on the test strip.

No, the Accu‑Chek Instant test strip does not offer any out-of-meter dosing option. To provide a timely, accurate result, electrochemical systems require a connection between the meter and test strip during the entire testing process.

Yes, as with any test strip, heat, humidity and other environmental conditions can affect strips.

No, the meters do not provide an automatic warning message telling customers that their vial of test strips will soon expire or lock them out once they have expired.

As with today’s products, the user will receive an E-4 error code.

No. There is no ability to squeeze extra blood from the finger to dose a strip a second time. However, there is a slight pause in testing that allows for the patient to take full use of the available blood from their finger.

The easy-edge of the Accu‑Chek Instant test strip has the largest dosing window of any major blood glucose test strip. It can be dosed anywhere along the edge and finger. Regardless of where dosed along the edge, the system provides a fast, accurate result.automatically absorbs the blood from the patient’s

The Accu‑Chek Instant test strip requires only a small speck of blood (0.6 μL).

Yes, a clear protective barrier runs along the entire length of the test strip to protect the strip and chemistry. Blood applied to the opening along the end of the strip can be quickly pulled into the test strip.

No, there is no compatibility between the Accu‑Chek Instant test strips with these and other Accu‑Chek systems.

No, the Accu‑Chek Instant and the Accu‑Chek Instant S meters do not require manual coding

Yes, it is possible to test with all types of blood including capillary, venous, arterial, and neonatal for all ROW countries outside the US and Canada.

The new Accu‑Chek Instant test strip utilizes the FAD-GDH chemistry. This system is the most accurate in the history of our organization. It meets the 10/10 standards for accuracy. Enzime FAD-GDH delivers advanced accuracy, highly reliable performance and reduced interferences.


Accu-Chek Instant

Item Description

Wishing there was an easy way to monitor your blood glucose levels? With the new Accu-Chek® Instant meter, it's as easy and fast as checking your phone messages. There is no setup required – just insert a test strip, apply a small blood sample and read your blood glucose result on the large display.

Product Manuals

WIRELESS TECHNOLOGY

Bluetooth® in Accu-Chek Instant

DISPLAY

Backlit LCD

RESULTS

Less than 4 seconds

SIZE

77.1 x 48.6 x 15.3 mm (LWH)

BATTERY TYPE

1 battery CR 2032 of 3V for Accu-Chek Instant S 2 batteries CR 2032 of 3V for Accu-Chek Instant (bluetooth connectivity feature)

TEST STRIP

Accu-Chek® The same test strip is used for Accu-Chek Instant and Accu-Chek Instant S meter.

LANCING DEVICE

Accu-Chek® Softclix

Benefits and Features
  • The intuitive target range indicator gives you visual reassurance. It can be individualised to suit your personal therapy goals
  • Only what you need: Your test result and your averages (7/30/90 days) are visible on the meter
  • No set up required – just insert a strip, apply a small blood sample, and read your blood glucose result on the large display
  • Easy-edge test strip: Apply a small drop of blood anywhere along the wide, yellow edge
  • Proven accuracy guaranteed by the makers of the Accu-Chek® brand, fulfilling the requirements of the ISO 15197:2013/EN ISO 15197:2015
  • The meter automatically logs your blood glucose results and wirelessly transfers them to your smartphone or via USB to your PC.
  • All data can be uploaded from the meter to your computer  via cable micro-USB. You are expected to install the Accu-Chek Smart Pix software in your computer.
  • Test strips come in a in a 50 vial.
  • Enjoy virtually pain-free testing with either the Accu-Chek® Softclix lancing device.
Additional specifications
Meter storage conditions:

Temperature: -25 – 70° C

Memory capacity:

The meter automatically stores at least 720 blood glucose results in memory, but only the last blood glucose result and your 7-, 30-, and 90-day averages can be viewed on the meter. To view stored blood glucose results, transfer them to a compatible software application.

Automatic off:

90 seconds after performing a test, 15 seconds after test strip is removed, or 5 seconds from the last test result screen.

Weight:

Approx. 40 g (with batteries)

Construction:

Hand-held

Protection class:

III

Interface:

USB; Bluetooth®

Radio frequency connectivity:

Bluetooth low energy technology operating in the frequency band of 2402 MHz to 2480 MHz with a maximum transmitted power of 0 dBm (1mW).